PRODUCT

EX Climatic Chamber

Standard Series

product name:EX Climatic Chamber

type:EX

function:

prize:

product detail


Model*1 EXT(H)250L EXT(H)340L EXT(H)600L EXT(H)1000L EXT(H)250U EXT(H)340U EXT(H)600U EXT(H)1000U
Power Supply AC 380V 50HZ 3 phase+N+PE
Maximum Current (EXT/EXTH) (A) 11 12.5 17 18.5/25.5 14 15 22 26/33
Maximum Power (EXT/EXTH) (KW) 5 5.3 7.5 11/12.5 6.8 7 10 15.5/17
Operation Environment +15~+35℃/≤85%RH
Control Mode Independent PID regulation of consecutive cooling and heating capacity
Performance*3 Temperature Range -40℃~+150℃ -70℃~+150℃
Relative Humidity Range +10℃~+90℃ 10%~98%RH
Dew Point Range +4℃~+88℃
Temperature Fluctuation ≤±0.5℃
Temperature Uniformity ≤2℃
Temperature Deviation ≤±2℃
Humidity Deviation ≤±3%RH
EXT Model Heating Rate*2 4.7℃/min 4.5℃/min 4.4℃/min 5℃/min 4.7℃/min 4.5℃/min 4.4℃/min 5℃/min
EXT Model Cooling Rate*2 3.2℃/min 3℃/min 3℃/min 4℃/min 3℃/min 2.8℃/min 3℃/min 5℃/min
EXTH Model Heating Rate*2 4.7℃/min 4.5℃/min 4.4℃/min 5℃/min 4.7℃/min 4.5℃/min 4.4℃/min 5℃/min
EXTH Model Cooling Rate*2 3℃/min 2.8℃/min 3℃/min 4℃/min 3℃/min 2.8℃/min 3℃/min 5℃/min
Inspection Window Multilayer glasses with conducting film heating, size W450mmxH450mm
Controller System AZ-1000 controller integrated with Siemens S7 PLC
Condenser Air-cooled condenser
Exterior Humidifying Water Supply Manually water supply to humidification water tank or self feeding by town water
Humidifying Water Recycling Multistage water purification device and water recycling, reducing water consumption
Interior Capacity (L) 225 337 573 1000 225 337 573 1000
Interior Size W(mm) 600 600 850 1000 600 600 850 1000
D(mm) 500 750 750 1000 500 750 750 1000
H(mm) 750 750 900 1000 750 750 900 1000
Exterior Size W(mm) 850 850 1100 1250 850 850 1100 1250
D(mm) 1460 1750 1750 2000 1460 1750 1750 2000
H(mm) 1925 1925 2075 2175 1925 1925 2075 2175
Standard Configuration ● Castors
● Cable port with silicon plug Φ80mmx1
● Internal lighting
● One shelf
● Air cooled condenser
● USB interface and history data review and converting software
Optional Configuration ○ Cable port with silicon plug Φ80mm/Φ150mm
○ Additional shelves
○ Water cooled condenser
○ Exterior socket 220V 10A
○ LAN interface and remote control software
○ Remote integration control for multi chambers
○ Digital alarm input channels x 4
○ Auxiliary relay digital output channels x 4 (for specimens)
○ Additional analog sampling channels x 4, sampling data can be reviewed and recorded (Option: PT100/T thermocouple/4~20mA/0~10V/-10V~10V)
○ Additional analog sampling channels x 8, sampling data can be reviewed and recorded (Option: PT100/T thermocouple/4~20mA/0~10V/-10V~10V)
○ Analog output channels x 4 (Setting temperature/Setting humidity/measured temperature/measured humidity)
○ Exterior analog input channels control temperature and humidity (Option: 4~20mA,0~10V, -10V~10V)
○ Compressed dry air purging system
○ Temperature control on test specimen
○ Short message monitoring system
*1: EXT modle has no humidity control function;
*2: Average speed between highest and lowest temperature(without load and heat dissipation according to IEC60068-3-5);
*3: All the above specifications are measured at +25℃ room temperature and the sensors are installed at blow-out of air conditioning unit;