PRODUCT

Speedy Climatic Chamber

Speedy,up to 20K/min

product name:Speedy Climatic Chamber

type:EX Speedy

function:

prize:

product detail

Model*1 EXT(H)600U-10-ESS EXT(H)600U-15-ESS EXT(H)1000U-10-ESS EXT(H)1000U-15-ESS EXT(H)2000U-10-ESS EXT(H)2000U-15-ESS
Power Supply AC 380V 50HZ 3 phase+N+PE
Maximum Current (EXT/EXTH) (A) 40 60 80 80 80 100
Maximum Power (EXT/EXTH) (KW) 22 32 40 40 40 54
Cooling water flow (m3/h) 5 6.5 9 9 9 10
Operation Environment +15~+35℃/≤85%RH
Control Mode Independent PID regulation of consecutive cooling and heating capacity
Performance*3 Temperature Range -70℃~+150℃
Relative Humidity Range +10℃~+90℃ 10%~98%RH
Dew Point Range +4℃~+88℃
Temperature Fluctuation ≤±0.5℃
Temperature Uniformity ≤2℃
Temperature Deviation ≤±2℃
Humidity Deviation ≤±3%RH
EXT Model Heating Rate*2 10℃/min 15℃/min 10℃/min 15℃/min 10℃/min 15℃/min
EXT Model Cooling Rate*2 10℃/min 15℃/min 10℃/min 15℃/min 10℃/min 15℃/min
EXTH Model Heating Rate*2 10℃/min 15℃/min 10℃/min 15℃/min 10℃/min 15℃/min
EXTH Model Cooling Rate*2 10℃/min 15℃/min 10℃/min 15℃/min 10℃/min 15℃/min
Inspection Window Multilayer glasses with conducting film heating, size W450mmxH450mm
Controller System AZ-1000 controller integrated with Siemens S7 PLC
Condenser Water-cooled condenser
Exterior Humidifying Water Supply Manually water supply to humidification water tank or self-feeding by town water
Humidifying Water Recycling Multistage water purification device and water recycling, reducing water consumption
Interior Capacity (L) 600 600 1000 1000 2000 2000
Interior Size W(mm) 850 850 1000 1000 1100 1100
D(mm) 750 750 1000 1000 1200 1200
H(mm) 900 900 1000 1000 1500 1500
Exterior Size W(mm) 1100 1100 1250 1250 1400 1400
D(mm) 2000 2000 2250 2250 2900 2900
H(mm) 2075 2075 2175 2175 2170 2170
Standard Configuration ● Castors
● Cable port with silicon plug Φ80mmx1
● Internal lighting
● One shelf
● Air cooled condenser
● USB interface and history data review and converting software
Optional Configuration ○ Cable port with silicon plug Φ80mm/Φ150mm
○ Additional shelves
○ Water cooled condenser
○ Exterior socket 220V 10A
○ LAN interface and remote control software
○ Remote integration control for multi chambers
○ Digital alarm input channels x 4
○ Auxiliary relay digital output channels x 4 (for specimens)
○ Additional analog sampling channels x 4, sampling data can be reviewed and recorded (Option: PT100/T thermocouple/4~20mA/0~10V/-10V~10V)
○ Additional analog sampling channels x 8, sampling data can be reviewed and recorded (Option: PT100/T thermocouple/4~20mA/0~10V/-10V~10V)
○ Analog output channels x 4 (Setting temperature/Setting humidity/measured temperature/measured humidity)
○ Exterior analog input channels control temperature and humidity (Option: 4~20mA,0~10V, -10V~10V)
○ Compressed dry air purging system
○ Temperature control on test specimen
○ Short message monitoring system
*1: EXT modle has no humidity control function;
*2: Average speed between highest and lowest temperature(without load and heat dissipation according to IEC60068-3-5);
*3: All the above specifications are measured at +25℃ room temperature and the sensors are installed at blow-out of air conditioning unit;